|
| ||||
|
CarboEurope R.E. Project Scientific Objectives Workpackage description Experiment plan 2005 Campaign 2007 Campaign CarboEurope R.E. data Information Upload site 2005 data access 2007 data access Partners Working papers News Join the mailing list Contact us ![]() |
The CARIBOU system (M.Ramonet, Ph. Galdemard)
CARIBOU instrument, a Non Dispersive Infra-Red spectrometer developed at CEA will be installed on a high tower for the intensive campaign of spring 2005. The system is divided in two subsystems:
figure VI.4.1-a: CARIBOU instrument : Pumping system (left), regulation & measurements system (middle) , Analysis and Control compartment (right) An overview of the Caribou design can be found below. The lower part of the figure shows the analysis compartment, which will includes a Peltier cooler to achieve an adequate dryness of the incoming air (the Peltier system is at the moment under manufacturing and could be replaced by a commercial cryocooler if not available). The pressure of the reference and sample cells are regulated individually and absolutely around 1080 mbar, and the test results show a relative pressure difference between them lower than 0.1 mbar. The flows of both the reference and sample gas are regulated at 20 cm3/min, which will allow very long lifetimes for the standard tanks (from half a year for the reference gas to several years for the other standards) CARIBOU site requirements:
figure VI.4.1-b: CARIBOU CO2 analyser schematics The CARIBOU CO2 analyser is based on a commercial LICOR-6252 CO2 and H2O analyser, which uses the differential, Non Dispersive Infra Red (NDIR) technique. The CO2 spectrum includes a number of absorption features in the thermal Infra Red. The absorption band centered at 4,26 µm is chosen in the LICOR-6252 because this region of the spectrum is quite clear from other features of other atmospheric species. figure VI.4.1-c: Principle of the LI-6252 differential, non-dispersive (NDIR) CO2 analyser figure VI.4.1-c depicts the operation principle of the LI-6252: The CO2 measurement is
inferred by measuring the difference in absorption of the infrared radiation
passing through two gas sampling cells. The reference cell is continuously
flushed with a gas of known CO2 concentration, and the sample cell is flushed
by the air sample or standard gas to be measured. The rotating chopping shutter
allows blocking alternatively the radiation entering the reference and the
sample cell, so the single IR PbSe detector alternatively measures the IR
intensity coming from the reference and the sample cells. The analyser output
is simply proportional to the difference in absorption between both cells.
After adequate calibrations of the instrument by known standard gases, the
analyser output is converted in CO2 absolute concentrations. The required level
of precision for the measurement of the CO2 concentration in the atmosphere, is
lower than 0.1 ppm for the CARIBOU instrument. This
is far better than what can be achieved using only a commercial LI-6252 from
the shelf. To increase the level of precision, the CARIBOU includes the
regulation of the physical parameters of the reference and sample gas: regulation of the LICOR cell pressure at 1080
mbar +/- 0.05 mbar regulation of the LICOR cells at 35°C +/- 0.03
°C. Finally, in order to minimise the running cost
of the CO2 station, both reference and sample gas flows are regulated at 20
cm3/min. The typical repetability of the CARIBOU
instrument is approximately 0.02 ppm rms (Note: this number is preliminary
because CARIBOU is still being characterised in the lab). See figure VI.4.1-d. figure VI.4.1-d: Typical repeatability of the CARIBOU analyser (preliminary results). The white
curve corresponds to the raw, uncalibrated measurements of the same standard
gaz, during 3 days. The orange curve is obtained by subtracting the varying
offset measured regularly by feeding the sample cell with the reference gas.
The standard deviation after offset correction is of the order of 0.02 ppm
|
||||
| Home> Experiment> Instrumental/technical details> CO2 concentration> LSCE (Biscarrosse)> | |||||
| top | © 2004-2007 Medias-France | ||||